Solutio in silico

TG-51 Practice Problems

A medical physicist is performing absolute dose calibration for a 6 MeV electron beam. The depth of 50% absorbed dose, R50, was measured and a value of 2.55 cm was obtained. All measurements were taken with a NE2505/3,3A cylindrical model ion chamber. The gradient correction factor was measured and a value of 0.9520 was obtained. The ion chamber-electrometer unit was calibrated to measure 4.994e+7 Gy/C under labratory conditions. The temperature in the clinic at the time of measurement was 20.8oC and the pressure was 767.2 mmHg. The average electrometer readings at bias voltages of 150 V, -150 V, and -300 V are 12.3330 nC, -12.3241 nC, and -12.3491 nC, respectively. Each measurement was taken after 100 machine monitor units (MU) had been delivered. What is the value of the absorbed dose to water at dmax for an SSD setup, in cGy/MU? (The PDD for this beam at clinical setup is 99.85%)

**All values, unless stated within the problem, should be obtained using the TG-51 report.

cGy/MU